RT Journal Article SR Electronic T1 A Comparison of INRs after Local Calibration of Thromboplastin International Sensitivity Indexes JF American Society for Clinical Laboratory Science JO Clin Lab Sci FD American Society of Chemistry and Laboratory Science SP 91 OP 95 DO 10.29074/ascls.15.2.91 VO 15 IS 2 A1 McGlasson, David L YR 2002 UL http://hwmaint.clsjournal.ascls.org/content/15/2/91.abstract AB There are approximately 300 reagent/instrument combinations for performing prothrombin times/international normalized ratios (PT/INR) in the United States. Manufacturers and laboratories continually struggle to ensure that the International Sensitivity Index (ISI) of their thromboplastin is accurate for assaying PT/INR.OBJECTIVE: This study reports the feasibility of a new method to locally calibrate ISI of thromboplastin using the mechanical STA automated coagulation analyzer (Diagnostica-Stago Inc.) and two photo-optic coagulation analyzers, the BCS (Dade-Behring) and CA-540 (Sysmex).DESIGN: Neoplastine CI+ (CI+) (Diagnostica-Stago Inc); Thromboplastin C+ (TC+); Thromborel S (TRS); and Innovin (I) (Dade-Behring) were used in this study. A mean normal PT (MNPT) was determined for each reagent/instrument combination using samples from 25 normal individuals. Manufacturer instrument specific ISI values were not available for the STA with TC+, TRS and I. The CA540 had no ISI value for CI+ and the BCS system had no manufacturer assigned ISI values for TC+ and I; generic photo-optic and mechanical ISI manufacturer values were used for these two systems. Local on-site calibration was performed using frozen plasma calibrators to determine ISI values for each thromboplastin. Post-calibration, 95 patient samples were assayed for each reagent/instrument system combination using the manufacturer ISI and the local calibrated ISI to determine the INR result.PATIENTS: Patients from whom samples were obtained included five with a lupus anticoagulant, 30 on heparin therapy, and 60 on coumadin therapy.RESULTS: Differences between manufacturer versus local calibrated ISI ranged from 0.9% to 18.9% for normal sample INRs and from 0.8% to 16.4% for patient sample INRs. The number (or proportion) of patient specimens with clinically significantly different INR values (>10.0% difference) ranged from zero for several reagent combinations to more than half (or >50.0%) of those tested for several other combinations.CONCLUSION: Our results indicated that by locally calibrating ISI values, each laboratory may eliminate variability and guesswork between different reagent/instrument systems for ISI values when performing PT/INR assays and potentially improve the clinical accuracy of their patients' PT/INR results.ABBREVIATIONS: CI+ = Neoplastine CI+; INR = international normalized ratio; IRP = international reference preparation; ISI = international sensitivity index; MNPT = mean normal prothrombin time; OAT = oral anticoagulant therapy; PIVKAS = proteins induced by Vitamin K antagonists; PT = prothrombin time; TC+ = Thromboplastin C+; TRS = Thromborel S; WHO = World Health Organization.