Table 1.

SARS-CoV-2 sequencing clinical quality metrics for use on the Ion Torrent Genexus

Quality MetricsFailure CutoffNeeds Review CutoffPass Cutoff
Run metricsLoading (%)<80N/A≥80
Enrichment (%)<95≥95
Library reads (%)<80≥80
Final reads (%)<30≥30
Final reads/lane<8 000 000≥8 000 000
Mean raw accuracy (%)<98≥98
Sample metricsTotal reads<100 000100 000–400 000>400 000
Mean read length<155≥155–174≥175
Uniformity of base coverage (%)<80≥80–<90≥90
Average base coverage depth<2000N/A≥2000
Percent reads on target (%)<90≥90–<95≥95
Number of mapped reads<50 000N/A≥50 000
Target base coverage at 20× (%)<80≥80–<90≥90

The metrics for quality control of SARS-CoV-2 sequencing on the Genexus can be divided into run-specific metrics and sample-oriented metrics. Most run metrics listed have hard cutoffs, while the sample metrics have a review window. Sequence data that “Pass” but need review should take all listed metrics into consideration and reference the qualitative metrics prior to reporting. NA, not available.